Transform absorbance data into transmittance values fast. Analyze solutions, compare optical readings, and export clean chemistry results with confidence daily.
This chart shows how absorbance changes transmittance on a logarithmic optical scale.
| Absorbance (A) | Decimal Transmittance (T) | Percent Transmittance (%T) | Fraction Absorbed | Interpretation |
|---|---|---|---|---|
| 0.1000 | 0.7943 | 79.43 | 0.2057 | Weak absorption |
| 0.3000 | 0.5012 | 50.12 | 0.4988 | Moderate absorption |
| 0.5230 | 0.3000 | 30.00 | 0.7000 | Strong absorption |
| 1.0000 | 0.1000 | 10.00 | 0.9000 | Very strong absorption |
| 2.0000 | 0.0100 | 1.00 | 0.9900 | Near-complete blocking |
1) From absorbance to transmittance:
T = 10-A
2) From decimal transmittance to absorbance:
A = -log10(T)
3) From percent transmittance to absorbance:
A = -log10(%T / 100)
4) From intensity values:
T = I / I0, then A = -log10(T)
5) Beer–Lambert relation:
A = εbc
Here, A is absorbance, T is decimal transmittance, %T is percent transmittance, I is transmitted intensity, I0 is incident intensity, ε is molar absorptivity, b is path length, and c is concentration.
Absorbance measures how much light a sample removes from an incoming beam. Higher absorbance means less light passes through the solution.
Transmittance is the fraction of light that passes through a sample. It may be shown as a decimal ratio or as percent transmittance.
Absorbance and transmittance are logarithmically linked. As a sample absorbs more light, less transmitted light remains, so transmittance drops quickly.
No. In standard measurements, percent transmittance should stay between greater than 0 and 100. Values above 100 usually suggest calibration issues.
Use the intensity method when you directly know sample intensity and reference intensity. The calculator converts their ratio into transmittance and absorbance.
Beer–Lambert mode estimates absorbance from molar absorptivity, path length, and concentration. It then converts the absorbance into transmittance values automatically.
In many lab contexts, optical density and absorbance are treated the same. Both describe logarithmic light loss through the sample.
Many instruments perform best within moderate absorbance ranges, often around 0.1 to 1.0. Very high values can reduce measurement reliability.
Important Note: All the Calculators listed in this site are for educational purpose only and we do not guarentee the accuracy of results. Please do consult with other sources as well.