Calculator inputs
Use consistent concentration units across standards and samples. This template reports concentration in mg/L and mass-normalized output in mg/g.
Example data table
| Field | Example value | Unit | Purpose |
|---|---|---|---|
| Standard analyte concentration | 25.0000 | mg/L | Known analyte concentration for response factor building |
| Standard internal standard concentration | 10.0000 | mg/L | Known internal standard concentration for normalization |
| Standard analyte peak area | 245,000 | area units | Measured analyte detector response for the standard |
| Standard internal standard peak area | 198,000 | area units | Measured internal standard detector response for the standard |
| Sample analyte peak area | 182,000 | area units | Measured analyte response for the unknown sample |
| Sample internal standard peak area | 204,000 | area units | Measured internal standard response for the unknown sample |
| Calculated RRF | 0.492462 | RRF | Relative response factor from the standard mixture |
| Corrected original concentration | 93.4119 | mg/L | Final analyte concentration after all corrections |
Formula used
CA,std,corr = CA,std × (PurityA / 100)
CIS,std,corr = CIS,std × (PurityIS / 100)
RRF = (AA,std / CA,std,corr) ÷ (AIS,std / CIS,std,corr)
CA,sample,prepared = (AA,sample / AIS,sample) × (CIS,sample,corr / RRF)
CA,sample,original = CA,sample,prepared × Dilution Factor ÷ Recovery Fraction
Amount (mg) = CA,sample,original × Final Volume (L)
Result (mg/g) = Amount (mg) ÷ Sample Mass (g)
How to use this calculator
- Enter analyte and internal standard names for clear reporting.
- Fill in the known standard concentrations and their measured peak areas.
- Enter the sample peak areas and the internal standard concentration added to the sample.
- Apply purity, dilution, recovery, final volume, and sample mass corrections.
- Submit the form to view response factors, corrected concentrations, export files, and the Plotly graph.
Frequently asked questions
1. What does an internal standard do?
It compensates for injection variability, detector drift, sample preparation loss, and matrix effects by normalizing analyte response against a stable reference signal.
2. Why are purity fields included?
Reference materials are rarely exactly 100 percent pure. Purity correction improves the response factor and prevents biased concentration estimates.
3. When should I use recovery correction?
Use recovery when your method validation shows incomplete extraction or transfer. A validated recovery factor can correct systematic underestimation.
4. Why does the calculator ask for final volume?
Final volume converts concentration into analyte mass. That allows reporting total analyte recovered in the prepared solution.
5. What unit system should I use?
Use one consistent concentration system across standards and samples. The page labels outputs as mg/L and mg/g for convenience.
6. Can I use this for GC, HPLC, or LC-MS?
Yes. The internal standard principle is the same when peak areas are integrated consistently and calibration assumptions remain valid.
7. What does normalized response mean?
It compares the sample area ratio with the standard area ratio. Values below 100 percent indicate weaker sample response relative to the standard.
8. Can one standard give accurate results?
A single standard can work for routine checks, but multi-level calibration usually provides stronger quantitation across wider concentration ranges.