Verify detector response using known gas and acceptance limits for worksites today. Record readings, flag failures, and schedule corrective calibration with reminders after testing.
| Date | Model | Gas | Target | After | Deviation | Resp (s) | Status |
|---|---|---|---|---|---|---|---|
| 2026-01-10 | Multi-gas X200 | H2S | 25 ppm | 24.6 ppm | -1.60% | 18 | PASS |
| 2026-01-12 | Multi-gas X200 | CO | 50 ppm | 56.5 ppm | +13.00% | 24 | FAIL |
| 2026-01-14 | Single-gas Pro | O2 | 18.0 %Vol | 18.1 %Vol | +0.56% | 12 | PASS |
This calculator compares the detector’s displayed reading against the test gas target concentration. A bump test is typically considered acceptable when the sensor responds and the reading is within a defined tolerance.
A practical field reference aligned with typical site safety programs.
Construction work often involves confined spaces, hot work, coatings, fuels, and changing ventilation. A bump test is a quick functional check that verifies a detector responds to a known gas challenge and that alarms activate. It supports daily readiness, especially before entry permits, tank work, sewer tie-ins, or demolition near unknown utilities.
A bump test confirms sensor response, alarm function, and basic flow path integrity. It does not replace calibration. Think of it as a “go/no-go” check: the display should rise toward the target concentration and alarms should trigger as expected. The calculator records target versus measured readings and highlights out-of-range results.
Many programs use a tolerance band such as ±10% to ±20% of the cylinder target, depending on sensor type and risk level. For example, a 50 ppm CO bump with a ±10% policy would accept readings from 45 to 55 ppm. This tool computes deviation percentage and compares it to your selected tolerance.
Response time limits are often set to confirm alarms occur promptly, such as within 30 seconds for a standard bump exposure. Longer times may indicate blocked inlets, depleted filters, weak pumps, or aging sensors. By entering observed response time and a maximum limit, the calculator flags slow response conditions.
Consistent delivery improves repeatability. Typical demand-flow or fixed-flow regulators deliver around 0.5 L/min, but always follow the detector and regulator instructions for the correct range. Keep caps on cylinders, avoid heat exposure, and check the cylinder expiration date before testing. Expired gas can create false failures or misleading passes.
Calibration intervals vary by policy, environment, and manufacturer guidance; 180 days is a common administrative interval. Harsh conditions, high humidity, solvent exposure, or frequent alarms may require more frequent calibration. This calculator identifies when the test date is beyond the last calibration date plus your interval, and recommends action.
Failures typically stem from incorrect gas type, wrong regulator, leaks in tubing, clogged inlets, depleted sensors, or expired cylinders. Corrective steps include confirming cylinder concentration and gas type, replacing filters, checking pump operation, and performing calibration. After corrections, retest and store exported records with permit or daily inspection logs.
Many sites bump test daily before first use, while lower-risk applications may use weekly schedules. Pick a frequency that matches hazard exposure, then document results consistently. The built-in next-bump date helps teams plan routine checks, reduce missed inspections, and strengthen audit readiness. Reliable documentation also supports trend review for sensors drifting toward failure.
No. A bump test checks response and alarms using a known gas challenge. Calibration adjusts the instrument to match a standard and usually requires a defined procedure and equipment.
Use your site or manufacturer policy. Many programs use ±10% to ±20% depending on sensor type and risk. Set a tighter band for higher-risk entries and critical alarms.
Treat it as a failure for readiness. Slow response can indicate blocked inlets, weak pumps, contaminated sensors, or filter issues. Inspect, service, and retest before relying on the unit.
Yes. Expired gas may have degraded concentration and can misrepresent sensor performance. Replace the cylinder, then repeat the bump test for a defensible record.
Many field checks use 20–30 seconds, but follow the detector guidance. Enter your actual bump duration to keep documentation consistent across shifts and crews.
Yes. Run a separate entry for each gas channel or cylinder challenge, keeping the same device model and test date. Store exports together as a single daily record.
Keep the test date, model, gas type, target, after-reading, deviation, response time, cylinder expiry, and calibration status. The CSV and PDF exports provide a consistent format.
Important Note: All the Calculators listed in this site are for educational purpose only and we do not guarentee the accuracy of results. Please do consult with other sources as well.