Compute OD using intensity ratios, percent transmittance, or Beer–Lambert law quickly today. Switch methods, review formulas, and download clean summaries for labs and classes too.
| Scenario | I₀ (a.u.) | I (a.u.) | T (%) | OD | Comment |
|---|---|---|---|---|---|
| Clear reference | 1.00 | 0.90 | 90 | 0.046 | Low attenuation, high transmission |
| Moderate absorber | 1.00 | 0.25 | 25 | 0.602 | Typical measurable range for many assays |
| Strong absorber | 1.00 | 0.01 | 1 | 2.000 | May require dilution to avoid saturation |
Optical density (OD) is a logarithmic measure of light attenuation through a sample. In spectrophotometry, OD equals absorbance, a dimensionless quantity that compresses large transmission changes into a manageable scale. An OD of 1 means 10% transmission; OD 2 means 1% transmission.
When your instrument reports incident and transmitted intensity, OD is computed from the ratio I/I₀. This calculator accepts any consistent intensity unit (counts, volts, or power). The ratio cancels units, but stable I₀ and low stray light are essential for trustworthy OD at higher values.
If you have transmittance directly, the relationship is straightforward: OD = −log₁₀(T). For example, 25% transmission (T = 0.25) corresponds to OD ≈ 0.602. Because T is bounded between 0 and 1, small measurement errors at low T can create larger OD uncertainty.
The Beer–Lambert model links absorbance to chemistry and geometry: A = εlc. Here ε is molar absorptivity, l is path length, and c is concentration. In many aqueous assays, l is 1 cm, while ε can vary widely by chromophore and wavelength, often from hundreds to tens of thousands L·mol⁻¹·cm⁻¹.
Most measurements are most reliable in moderate absorbance ranges where detectors remain linear. A common working window is roughly OD 0.1 to 1.0, though this varies by instrument. Very low OD can be dominated by noise; very high OD can be dominated by stray light and baseline drift.
Blank OD accounts for cuvette, solvent, and background absorption. Subtracting a measured blank improves comparability across runs and helps reduce systematic bias. This calculator applies blank correction consistently across all methods, including Beer–Lambert computations, so your reported OD aligns with analytical practice.
High OD samples are commonly diluted to return to a linear range. The dilution factor scales concentration back to the original sample. For nonstandard cuvettes or microplates, path length differs from 1 cm; adjusting l in Beer–Lambert calculations preserves accuracy. Always keep concentration and ε units compatible.
Good reporting includes method, OD, transmittance, wavelength, path length, and any dilution used. The built-in CSV export is useful for spreadsheets and lab notebooks, while the PDF export provides a quick one-page summary for documentation. Consistent metadata helps with audit trails and reproducibility.
In most spectrophotometry contexts, yes. OD is commonly used as another name for absorbance A, defined as −log₁₀(T). Some fields use OD for filter density, but the math is equivalent.
OD = 0 means T = 1, or 100% transmission through the sample relative to the reference. In practice, tiny offsets may exist due to noise and baseline effects, so blanks help.
At high absorbance, transmitted light is very small, so stray light, detector limits, and baseline drift can dominate. That can make the computed OD appear lower than reality. Dilution or shorter path length improves reliability.
Use intensity ratio if you have I₀ and I readings, transmittance if your device reports %T or T directly, and Beer–Lambert when you need chemistry-based conversions such as concentration, ε, or path length.
Enter the OD of a blank sample measured under the same conditions (same cuvette, solvent, wavelength, and instrument settings). If you do not have a blank measurement, leave it at zero.
If you diluted the sample before measuring, enter the dilution factor (for example, 10 for a 1:10 dilution). The calculator reports concentration scaled back to the original sample when solving for concentration.
Negative OD can occur if the sample transmits slightly more than the reference due to noise or imperfect blanking. Re-check baseline, cuvette cleanliness, and reference measurements; then remeasure or apply proper blank correction.
Accurate optical density guides measurements, safety, and quality control.